Set Inst Time
Use á, â, and ENTER (8 ) to select YES.
•
•
Test Stand will update instrument time on each insertion or
Use á, â, and ENTER (8 ) to select NO.
•
•
Test Stand will not alter instrument time on each insertion.
Buzzer
Use á, â, and ENTER (8 ) to select ON or OFF for the Test
•
Stand buzzer.
Set Password
Use á, â, and ENTER (8 ) to select ON, OFF, or CHANGE for the
•
Test Stand setup password.
After selecting CHANGE, use á, â, and ENTER (8 ) to select the
•
new desired password.
•
672 is always accepted and allows entry to select a new password.
Power Save (automatically enabled when using a battery pack)
•
Power save mode will turn OFF the Test Stand after it is idle for
one minute (or will beep every 15 seconds if an instrument is left in
the Test Stand).
Use á, â, and ENTER (8 ) to select ON or OFF for the Test Stand
•
power save mode.
Inst Cal Mode
•
Instrument cal mode determines whether or not the instrument
stays in calibration mode after a calibration is finished.
•
By default, instrument calibration is enabled; this mode prevents
the instrument from alarming while on the Test Stand.
Use á, â, and ENTER (8 ) to select ON or OFF to enable or
•
disable the instrument cal mode.
Internal DB
•
Each Galaxy Test Stand has enough internal memory to store a
database of up to 150 instruments (an error message will signal
when this limit is hit).
•
To prevent this error message, use the memory card option
and disable the internal database.
3-4