Test Setup; Gas Setup - MSA Galaxy Manuel D'instructions

Système automatisé de tests
Masquer les pouces Voir aussi pour Galaxy:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Use á, â, and ENTER (8 ) to select ON or OFF to enable or
disable the internal database.

Test Setup

Use á, â, and ENTER (8 ) to select from the following Test Stand
operating modes:
Bump Only
Test Stand only BUMPs and reports a pass or fail status.
Cal Only
Test Stand always performs a full calibration on the
instrument.
Bump Fail Cal
Test Stand BUMPs.
If the instrument fails BUMP, the Test Stand performs a full
calibration.
Bump Time
Select time (in seconds) for bump tests.
Default is 60 seconds.
Bump Limits
Select limits (in %) for bump tests.
Default is +10%.

Gas Setup

In Gas Setup:
The user can change the gases assigned to each cylinder
attached to the Test Stand.
Standard Test Stands have limited cylinder options and are
designed for using QuadGas cylinders.
Standard Solaris Test Stand allows FAS and one cylinder.
Standard Sirius Test Stand allows FAS and two cylinders.
Test Stands that allow up to FAS and three cylinders can be
ordered.
If the Test Stand is NOT using cylinder holders, the gas
connections on the left-most Test Stand are as follows from top:
Fresh Air
3-5

Publicité

Table des Matières
loading

Table des Matières