Use á, â, and ENTER (8 ) to select ON or OFF to enable or
•
disable the internal database.
Test Setup
Use á, â, and ENTER (8 ) to select from the following Test Stand
•
operating modes:
•
Bump Only
•
Test Stand only BUMPs and reports a pass or fail status.
•
Cal Only
•
Test Stand always performs a full calibration on the
instrument.
•
Bump Fail Cal
•
Test Stand BUMPs.
•
If the instrument fails BUMP, the Test Stand performs a full
calibration.
•
Bump Time
•
Select time (in seconds) for bump tests.
Default is 60 seconds.
•
Bump Limits
•
Select limits (in %) for bump tests.
Default is +10%.
Gas Setup
In Gas Setup:
•
The user can change the gases assigned to each cylinder
attached to the Test Stand.
•
Standard Test Stands have limited cylinder options and are
designed for using QuadGas cylinders.
•
Standard Solaris Test Stand allows FAS and one cylinder.
•
Standard Sirius Test Stand allows FAS and two cylinders.
•
Test Stands that allow up to FAS and three cylinders can be
ordered.
•
If the Test Stand is NOT using cylinder holders, the gas
connections on the left-most Test Stand are as follows from top:
•
Fresh Air