6. OPERATION
6.1 Making the scale operational
Moving the scale
Aligning the scale
2 .
82 •
CAUTION!
Bodily injury
Perform a function check as described in the
section entitled "Function check" on page 102
before using the device each time.
1. Tilt the device until it can be moved freely on the
casters.
2. Move the device in this position to the desired set-
up or storage location.
ATTENTION!
Incorrect measurement due to force shunts
If the scale and housing is in contact with some-
thing, e.g. a towel, weight will not be measured
correctly.
– Set the scale up so that only its foot screws
1. Place the scale on firm, level surface.
2. Undo the knurled wheels.
3. Level the device by turning the foot screws.
4 .
The air bubble in the spirit level must be located in
the exact center of the circle.
3 .
are in contact with the floor.