MD 9070 Insulation/Continuity Digital Multimeter
~ Hz Line Level Frequency
Function RANGE Sensitivity (Sine RMS) Range
6 V
60 V
600 V
1000 V
VFD 600 V
Accuracy: 0.02%+4d.
1)
VFD sensitivity linearly decreases from 10% F.S. @ 200 Hz to 40% F.S. @ 440 Hz.
Insulation Resistance
Test Voltage
50 V
100 V
250 V
500 V
1000 V
1000 V
1000 V
1)
Actual output voltage: 100% ~ 120% of Test Voltage.
Live Circuit Detector: Inhibit test and display voltage reading instead if terminal voltage >
30 V prior to initialization of test. Display voltage accuracies:
DCV: 1,5% + 5d.
ACV: 3,0% + 5d @ 50 Hz ~ 60 Hz (unspecified @ > 600 Vac).
Specified measuring range is 0,020 MΩ ... 25,0 GΩ for percentage operating
uncertainty B[%] ≤ ±30% per IEC/EN61557-2 requirements.
0,6 V
6 V
60 V
600 V
60 V ~ 240 V
1)
Range
3,000 MΩ, 30,00 MΩ, 55,0
MΩ
3,000 MΩ, 30,00 MΩ, 110,0
MΩ
3,000 MΩ, 30,00 MΩ, 275,0
MΩ
3,000 MΩ, 30,00 MΩ, 300,0
MΩ, 550,0 MΩ
3,000 MΩ, 30,00 MΩ, 300,0
MΩ
3000 MΩ
25,0 GΩ
25
10 Hz ~ 20 kHz
10 Hz ~ 20 kHz
10 Hz ~ 3 kHz
10 Hz ~ 3 kHz
1)
10 Hz ~ 440 Hz
Test Current
1 mA @50 kΩ 1,5%+5d
1 mA @100 kΩ 1,5%+5d
1 mA @250 kΩ 1,5%+5d
1 mA @500 kΩ 1,5%+5d
1 mA @1 MΩ 1,5%+5d
1 mA @1 MΩ 2,0%+5d
1 mA @1 MΩ 10%+5d
Specifications
Accuracy