MD 9070 Insulation/Continuity Digital Multimeter
5 MAINTENANCE
WARNING
To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input jacks and turn OFF the meter before opening the case. Do not operate
with open case. Install only the same type of fuse or equivalent
Calibration
Accuracy is specified for a period of one year after calibration. Periodic calibration at
intervals of one year is recommended to maintain meter accuracy.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives
or solvents. If the meter is not to be used for periods of longer than 60 days, remove the
battery and store it separately
Trouble Shooting
If the instrument fails to operate, check battery, fuses, leads, etc., and replace as
necessary. Double check operating procedure as described in this user's manual
If the instrument voltage-resistance input terminal has subjected to high voltage
transient (caused by lightning or switching surge to the system) by accident or abnormal
conditions of operation, the series input protection resistors may be blown off (become
high impedance) like fuses to protect the user and the instrument. Most measuring
functions through this terminal will then be open circuit. The series input protection
resistors and the spark-gaps (or varistors) should then be replaced by qualified
technician. Refer to the LIMITED WARRANTY section for obtaining warranty or
repairing service.
Battery and Fuse replacement
Battery use: Four 1.5V AA battery (IEC LR6)
Fuses use:
Fuse (F2) for
0,4 A / 1000 Vac & Vdc, IR 30 kA or better, FF fuse; Dimension: 6 x 32 mm.
Fuse (F3) for Earth Continuity Test input:
0,25 A / 1000 Vac & Vdc, IR 30 kA or better, FF fuse; Dimension: 6 x 32 mm.
/mA input:
20
Maintenance