Reprocessed LigaSure Impact™ Curved, Large Jaw, Open Sealer/Dividers
1
Directions for Use
1. Before beginning the procedure, verify overall compatibility of all instruments and accessories.
2. Inspect packaging before opening. The contents of the package are sterile if the package has not been compromised. Do
not use the instrument if the sterility has been compromised. If the package is damaged or if it was opened and the
instrument not used, return the instrument and the package to Stryker.
3. Do not attempt to resterilize. Stryker will not accept instruments for reprocessing that have been reprocessed and
sterilized by other facilities.
4. Remove the instrument from the package by firmly pulling on the handle, taking care not to pull on the instrument's jaws
or connector cable. Place it in a sterile work are using aseptic technique.
5. Inspect the instrument for overall condition and physical integrity. Do not use the instrument if any damage is noted.
Return the instrument and packaging to Stryker if it is not in acceptable condition for the procedure.
6. Follow a suitable surgery protocol.
7. Instrument is intended for use during single patient procedure.
Set Up
Warning
Electrical Shock Hazard - Do not connect wet accessories to the LigaSure system.
Confirm proper energy platform settings before proceeding with surgery.
Inspect the instrument and cords for breaks, cracks, nicks, or other damage before use. Failure to observe this caution
may result in injury or electrical shock to the patient or surgical team, or cause damage to the instrument. If damaged, do
not use.
Position instrument cords to avoid contact with the patient or other leads. Do not wrap cords around metal objects. This
may induce currents that can lead to shocks, fires, or injury to the patient or surgical team.
Ensure all connections to the energy platform and all instruments and accessories are secure before using. Improper
connection may result in arcs, sparks, accessory malfunction, or unintended surgical effects.
Insert the connector into the receptacle on the generator. Follow the instructions in the generator user's guide to complete
the setup procedure.
LigasSure function During Surgery
Warning
Fire Hazard - Do not place instruments near or in contact with flammable materials (such as gauze or surgical drapes).
Instruments that are activated or hot from use may cause a fire. When not using instruments, place them in a clean, dry,
highly visible area not in contact with the patient. Inadvertent contact with the patient may result in burns.
Use the system with caution in the presence of internal or external pacemakers. Interference produced by the
electrosurgical equipment can cause a pacemaker to enter an asynchronous mode or block the pacemaker effect entirely.
Consult the pacemaker manufacturer or hospital cardiology department for further information when use in planned in
patients with cardiac pacemakers.
4
3
2
5
1
LigaSure Cable Connector
2
Rotation Wheel
3
Handle
4
Activation Button
5
Cutting Trigger
Page 3 of 16