Reagent cartridges are for single-use only. Do NOT refill. Any
refilling or reusing of reagent cartridges will void all Analyzer and
parts warranties and nullify any performance claims.
Keep the OPAQUE type reagent cartridge in an upright position
before attaching and detaching the reagent supply line to avoid
exposure to the chemical reagent. Save the plug to later reinsert
it into the nozzle inlet after removing the cartridge from the
Analyzer during a cartridge replacement process.
To protect against accidental exposure to ultra-violet radiation, do
not operate the UV lamp outside of its protective housing.
(On-Line and Portable TOC Analyzers) Water in the iOS System
may be hot. Before inserting a vial into the iOS System after
operating in On-Line mode, slide the door open and wait 30
seconds to allow sample to completely drain. Inserting a vial
before draining can result in hot water spray projecting upward
out of the iOS System.
The iOS System and vial ports contain sharp needles designed to
pierce the septa of sample vials. Do NOT put fingers or
inappropriate materials into the iOS System or vial port.
When servicing parts inside the Analyzer, ensure that power to
the Analyzer is OFF, and keep hands clear of the reagent syringe
assemblies. The syringes are controlled by moving parts that can
pinch skin.
(Laboratory and Portable TOC Analyzers) For continued
protection against fire hazard, replace fuse with same type and
rating.
To avoid potentially dangerous shock, do not touch anything
inside the Analyzer while observing the sample pump.
This procedure involves temporarily installing an Ethernet cable
from the back of the instrument into an internal Ethernet port and
should ONLY be performed by a qualified technician.
Analyseurs COT Sievers M9/M9
DLM 77020-01 FR Rev. A
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
W
ARNINGS
e
339 / 439
AVERTISSEMENTS
Manuel d'utilisation et de maintenance
SUEZ © 2020