The Analyzer requires an A/C mains supply of 100-240 volts and
requires the correct power cord for safe operation. ONLY use the
power cord supplied in the Analyzer's Accessory Kit with the
Analyzer, which meets this A/C mains supply requirement.
For safety reasons and to avoid the possibility of corrupting the
database, ALWAYS follow this procedure when shutting down the
Analyzer. First stop analysis AND stop the operating system
(using the Shut down button in the Analyzer's firmware) BEFORE
pressing the Analyzer's power switch to Off or unplugging the
power cord,
Any operation requiring access to the inside of the Analyzer,
including installation of maintenance items, could result in injury.
To avoid potentially dangerous shock do this. BEFORE opening
the Analyzer, stop any analysis and shut down the operating
system (using the Shut down
firmware), then turn off power and disconnect from the power
supply.
Electrical conduit and output and alarm connections should be
installed by a qualified electrician. ESD protection should also be
exercised.
This is a Class A product. In a domestic environment, this product
may
measures to correct the interference.
Hazardous reagents (ammonium persulfate and phosphoric acid)
are used in the Analyzer. Before installing a reagent, read the
corresponding Material Safety Data Sheet (MSDS) for proper
handling precautions and spill or leak procedures. The MSDS are
contained in the pouch located on the outside of the reagent
shipping box.
Installation of reagents and most Analyzer maintenance requires
access to the inside of the Analyzer. To avoid potentially
dangerous shock do this. BEFORE opening the Analyzer, stop
any analysis and shut down the operating system (using the Shut
Down
and disconnect from the power supply.
To avoid exposure to the chemical reagents, wear acid-resistant
gloves and safety goggles.