EN
15
Loading the rotor
Standard centrifuge containers of glass will not stand RCF values exceeding 4000 (DIN 58970, pg. 2).
•
Check the rotor for firm seating.
•
With swing-out rotors all rotor positions must be lined with identical hangers. Certain hangers are marked with
the number of the rotor position. These hangers may only be used in the respective rotor position. Hangers that
are marked with a set number (e.g. S001/4) may only be used in the set.
•
The rotors and hangers may only be loaded symmetrically. The centrifuge containers have to be distributed
evenly on all rotor positions. For authorised combinations see Chapter "Anhang/Appendix, Rotoren und
Zubehör/Rotors and accessories".
In the case of angle rotors all possible rotor positions must be loaded, see chapter "Anhang/Appendix, Rotoren
und Zubehör/Rotors and accessories".
Rotor is loaded symmetrically.
Rotor is evenly loaded
•
On certain hangers, the weight of the maximum load or the weight of the maximum load and the maximum
weight of the completely loaded hanger is specified. This weight may not be exceeded. In case of exception, see
chapter " Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm
specification of the maximum load includes the total weight of the adapter, centrifuge container and contents.
•
In containers with rubber inserts, the same number of rubber inserts must always be among the centrifuge
containers.
•
The centrifuge containers may only be filled outside of the centrifuge.
46/146
Not permitted!
Rotor is loaded asymmetrically.
Not permitted!
Rotor is not evenly loaded
Rev. 08 / 04.2020
3
". The weight
AB1401DEENFRIT