Schémas de circuits modifications par rapport à la dernière édition

Publicité

+
+
C2
, M1
RS232 Interface
Dental workstation → PC
An extended Mouse System Protocol is used to transmit defined conditions or commands from the dental workstation to a connected PC.
For the camera and foot switch conditions in byte 1, the bits 0-2 are used to control the video application, to transmit additional commands, the bits
3-6 are used. The bytes 2-5 are 0 and are not used.
7
6
5
4
3
2
1
User command
Video function
Byte 1 of the Mouse System
In each case only one user command or information to the video function can be transmitted in one data package. The bits of the other block are set
to 0.
The following events and interface commands are defined:
The commands are generated in each case when the event is triggered.
+
C2
Rhomb key activated
Rhomb deactivated
Call key
X-ray image viewer key
CCW rotation key
(with instruments deposited or
video camera removed)
Instrument light key
(with instruments deposited or
video camera removed)
Quick setting key 100
(with instruments deposited or video camera removed)
Quick setting key 75
(with instruments deposited or video camera removed)
Quick setting key 50
(with instruments deposited or video camera removed)
Quick setting key 25
(with instruments deposited or video camera removed)
Quick setting key 1
(with instruments deposited or video camera removed)
Forward cursor from chair key
block
Forward cursor from instrument
key block
Spray key
Left discrete C+ FS button active
with camera removed
Chip blower key
Right discrete C+ FS button
active with camera removed
Camera removed
Camera deposited
Foot switch active with camera removed
Foot switch inactive with camera removed
PC → Dental workstation
The RS232 configuration and service information can be output via PC with an 1-byte command.
Command
0DH
Data transmission rate:9,600 Baud
58 71 657
1
0
Event
+
M1
White screen videokey
Next image videokey
Full screen videokey
Clear videokey
Black screen videokey
Save videokey
Left discrete C+ FS button active
with camera removed
Quad image videokey
Right discrete C+ FS button
active with camera removed
Action
Output of configuration and
service information
Blatt
10A
Sheet
Feuille
Hoja
Remark
+
+
C2
M1
only if switching mode
also be
operated via FS
cannot be
operated via FS
cannot be
operated via FS
cannot be
operated via FS
also be
operated via FS
only be
operated via FS
only be
operated via FS
Remark
RS232 command 32
(1st byte of the 5-
byte protocol)
+
+
C2
M1
88H/136
90H/144
98H/152
A0H/160
A8H/168
B0H/176
B8H/184
E8H/232
C0H/192
C8H/200
D0H/208
D8H/216
E0H/224
E8H/232
B8H/184
F0H/240
F8H/248
81H/129
83H/131
85H/133
81H/129