The Finnpipette FOCUS Multichannel Plus is an autoclavable pipette. It operates on the air
displacement principle and uses detachable, disposable tips.
The adjusted delivery volume is displayed on a numeric readout window on the handle.
The seven different models of Finnpipette FOCUS Multichannel Plus pipettes cover a
volume range from 1 μl to 300 μl.
Order No.
Channel
4610100
4610110
12
4610120
4610130
12
4610140
4610150
12
4610170
16
Numeric display
The adjusted delivery volume is clearly indicated in the numeric display on the handle.
Raw materials
The Finnpipette FOCUS Multichannel Plus is made of mechanically durable and chemically
resistant materials which allow repeated autoclaving of the complete pipette at 121°C
(252°F).
The user should verity that the pipette and tip is resistant to the liquid which it is intended
to be used.
Description of tips
Finntips are recommended for use with the Finnpipette FOCUS Multichannel Plus.
They are made of natural colour polypropylene, generally regarded as the only contamination
free material suitable for tips. Finntips are also autoclavable (121°C / 252°F).
Setting the delivery volume
1. Rapid adjustment: Set the delivery volume using the
push button on the top of the pipette. To increase the
delivery volume, turn the push button counter clockwise.
To decrease the delivery volume, turn it clockwise.
2. Fine adjustment: Set the last digit (0-9) of the delivery
volume by turning the fine adjustment ring.
3. Make sure that the desired delivery volume clicks into
place and that the digits are completely visible in the
display window.
The adjustable volume is limited to the specific volume
range. Using excessive force to turn the push button outside
the range may jam the mechanism and eventually damage the pipette.
Tip ejection
To help eliminate the risk of contamination, each pipette is fitted with a tip ejector system.
The tip ejector system consists of a soft-touch tip ejector and specially designed gearing
mechanism. To release the tip, point the pipette at suitable waste receptacle and press the
tip ejector with your thumb.
Shelf hanger
You can attach the pipette shelf hanger on a counter, pipette stand or anywhere where you
want to hang your pipette.
Clean the area where you plan to attach the shelf hanger. Apply two stickers to the
underside of the shelf hanger. Press the shelf hanger firmly into place — on a shelf, counter
top or pipette stand. To use, hang the grippy finger rest on the shelf hanger.

Product description

Volume Range
8
1 μl
to
1 μl
to
8
5 μl
to
5 μl
to
8
30 μl
to
30 μl
to
5 μl
to

Pipette operation

4
Finntip
10 μl
10
10 μl
10
50 μl
250 Universal, 200 Ext
50 μl
250 Universal, 200 Ext
300 μl
300, Flex 300
300 μl
300, Flex 300
50 μl
50
1
2