Lubricants
In the following table are reported the oil recommended for generic
use.
Tab. 6
Ambient temperature
12 – 35 °C
In case of change of the type of lubricant, it is necessary to proceed
with a pump by letting it run at maximum vacuum for at least 5
minutes.
For this operation use a quantity of oil such as to reach the minimum
level indicated on the tank.
Discharge again the oil and proceed with new oil charge.
Decommissioning
Drain the oil from the pump prior to the removal. If the oil is
polluted, flush the pump with fresh oil (see "Oil change").
Drain the oil from the tank, plug the inlet and the discharge ports and
store the pump without oil.
MS120 Single Stage Rotary Vane Pumps (G2571-64000) User Manual/X3702-90000 (A) 159/188
Lubricants
Viscosity
ISO 53
Instructions for Use
Lubricants
Agilent Technologies oil
AVF 60 Gold
4