Télécharger Imprimer la page

Use; Expected Use; Forbidden Use - Agilent Technologies MS120 G2571-64000 Notice De Mode D'emploi

Publicité

Use

Expected use

These vacuum pumps have been designed to handle air and small
quantity of water vapor only. Handling of other types of gas or vapors
must be declared in advance to Agilent Technologies that will give the
conformity to the specific use.
They are suitable to evacuate closed systems or to operate at a
constant vacuum within the following vacuum range: 0.07 - 15 mbar
(abs.).
The ambient temperature and the inlet temperatures must be
included between 12° and 35 °C.

Forbidden use

WARNING!
The pump must not handle:
WARNING!
It is prohibited to use the discharge of the pump to create even limited
pressures
MS120 Single Stage Rotary Vane Pumps (G2571-64000) User Manual/X3702-90000 (A) 146/188
-
liquids or solid substances
-
dangerous, explosive or aggressive gases and vapors
-
pure oxygen or air mixtures enriched with oxygen.
Instructions for Use

Use

4

Publicité

loading