5
Technical Information
About Commands
User Operator Report Type 2 and User Technician
Report Type 3
Report Format:
"HVc sssss iiiiixA vvvvV F=ffff E=eeee pppppmW i=yyyy v=zzz
c=www <CR>"
Report Fields:
HVc
sssss
iiiii
vvvv
ffff
eeee
ppppp HV Channel Power in mW
yyyy
zzz
www
Example:
RT 2[Enter]
+
HV2 OFF
v=000 c=D80
HV1 OFF
v=000 c=D80
HV2 OFF
v=000 c=D80
94/106
HV Channel c= 1 or 2
HV Channel Status ['OFF__', 'ON___', 'FAULT']
HV Channel Current in nA or uA
HV Channel Voltage in V
Hexadecimal representation of HV Channel Faults
Hexadecimal representation of HV Channel Events
16 Bits hexadecimal Current
10 Bits hexadecimal Voltage
12 Bits hexadecimal Control Output
0uA
0uA
0uA
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)
0V F=0000 E=0000
0V F=0000 E=0000
0V F=0000 E=0000
0mW i=0001
0mW i=0003