ION Pump Controller Description
Output for the pumps:
With Line Supply:
Output Voltage (each channel)
Output Current (each channel)
Maximum Output Power (each
channel)
Current Recorder Output signal
Operating Temperature
Storage Temperature
Compliance to Norms
Line Cable
Line fuse
High Voltage Connection
All connecting cables for remote I/O and serial connections, if longer than 3 m,
NOTE
must be of the shielded type.
76/106
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)
Default value = +5000 V +1/-5 %, settable from
+3000 – +5000Vdc +1, -5 %, step +/- 50 V by
serial line command
>30 mA ± 10 %
40 W
0 to 5 Vdc logarithmic,
proportional to 1.0e-8 to 1.0e-1 A
1 recorder output for each pump.
Minimum recorder input impedance: 1MOhm
Precision: Better than 1%
Current Measurement Resolution: 10nA under
100uA; 1uA over 100uA.
5 °C to 40 °C
-20 °C to 70 °C
Radio Interference: EN55011 Class A Group 1
Harmonics:
Flickers:
ESD:
Burst:
Radiated RF Immunity:
Surge:
Conducted noise imm.:
Line supply interruptions:
Safety:
3 m long cable (3 wires), no plug
Two T5A
3 High Voltage Fischer connectors
EN61000/3/2
EN61000/3/3
EN61000/4/2
EN61000/4/4
EN61000/4/3
EN61000/4/5
EN61000/4/6
EN61000/4/11
EN61010-1