Report Type 0; User Operator Report Type 0 And User Technician Report Type - Agilent Technologies 629-5001 Notice De Mode D'emploi

Publicité

Report TYPE 0...6

The User Operator Reports Type 0, 2, 4 and 6 display the 'continuous
average value' for the HV Channel Current, Voltage and Power, both
in normal or scientific notation and the 'instantaneous measured or
control values' in the hexadecimal format.
Technical Reports Type 1, 3 and 5 report the 'instantaneous
measured values' for the HV Channel Current, Voltage and Power,
both in normal or scientific notation and the 'continuous average
measured values' in the hexadecimal format.
User Operator Report Type 0 and User Technician
Report Type 1
Report Format: "HVc sssss iiiiixA vvvvV F=ffff E=eeee <CR>"
Report Fields:
HVc
sssss
iiiii
xA
vvvv
ffff
eeee
Example:
RT 0[Enter]
+
HV2 OFF
HV1 OFF
HV2 OFF
Ion Pumps Control Unit User Manual / 87-900-124-01
HV Channel c= 1 or 2
HV Channel Status ['OFF__', 'ON___', 'FAULT']
HV Channel Current in xA
nA or uA (nano Ampere or micro Ampere)
HV Channel Voltage in V (Volts)
Hexadecimal representation of HV Channel Faults
Hexadecimal representation of HV Channel Events
0uA
0V F=0000 E=0000
0uA
0V F=0000 E=0000
0uA
0V F=0000 E=0000
Technical Information
(C)
5
PIN Functions
93/104