Technical Information
ION Pump Controller Description

Output Signals Connection

Relay outputs: Single contact
Output relays rating: 2 A @ 125 Vac.
Minimum load: 10 mW (100 mV @ 1 mA).
Bounucing Time: NO=1msec; NC=3msec.
Relay contact "Closed" when I>Iset-point
Relay contact "Open" when I<Iset-point
Relay Idle State: "Open" when unit in main power OFF condition
Relay contact "Closed" when HV = ON
Relay contact "Open" when HV = Off
Relay Idle State: "Open" when unit in main power OFF condition
Relay contact "Closed" (Relay deactivated) when unit in normal
operation
Relay contact "Open" (relay activated) when unit in normal operation
Relay Idle State: "Closed" when unit in main power OFF condition
80/106
Set-Point Relay Outputs:
Ch1 Set-Point: Pin 1 (NO), Pin 14 (COM)
Ch2 Set-Point: Pin 2 (NO), Pin 15 (COM)
CH3 Set-Point: Pin 3 (NO), Pin 16 (COM)
HV Status Relay Outputs
Ch1 HV status: Pin 4 (NO), Pin 17 (COM)
Ch2 HV Status: Pin 5 (NO), Pin 18 (COM)
Ch3 HV Status: Pin 6 (NO), Pin 19 (COM)
Fault status relay contact:
Fault status: Pin 7 (NC), Pin 20 (COM)
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)