Installation Procedure
Operating Modes
Operating Modes
Power Line Supply
LOCAL-REMOTE I/O Operation
The 3 HV switches on the front panel and the 3 HV On/Off inputs on the DB25
NOTE
"Remote" connector on the rear panel are in series.
The HV to the 3 pumps can be switched ON or OFF by the 3 HV (HV1,
HV2 or HV3) switches on the front panel (in this case the respective
HV ON-OFF signal on DB25 remote connector must be closed and the
respective HV Cable Interlock must be closed) or can be switched ON
or OFF by short-circuiting the HV on-off signal on the remote
connector (on the rear panel) with the related GND reference (in this
case the respective HV switch on the front panel must be in the ON
position and the respective HV Cable Interlock must be closed).
When the HV is ON, the corresponding HV LED on the front panel,
will be ON.
If the connection between the IPCU and the pump is done by means of an HV
NOTE
Cable with Safety Interlock, the cable must be connected to the pump F/T and
the small banana connector on the cable controller side must be inserted in the
corresponding "HV Cable Interlock" plug on the unit rear panel in order to be
able to switch on the HV.
If the connection is done with a single pole HV Cable without the
Safety Interlock, ensure that the jumper provided on the unit rear
panel is inserted in the "HV Cable Interlock" plug.
68/106
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)