Usage
This paragraph gives a general information on the use of the
controller. For details refer to the "Technical Information"
paragraph. When the controller is plugged to the Power Line and the
Line Switch on the rear panel is ON (I), it is able to operate with its
full set of features:
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)
3 independent HV channels
+ 5000 Vdc maximum output Voltage
40 W maximum output Power for each channel
> 30 mA maximum output current (short circuit condition) for
each channel
LOCAL-REMOTE I/O/SERIAL operating mode
START/PROTECT operation
Installation Procedure
Usage
4