Figure 7

Input Commands Connection

Input signals:Optically Insulated
Insulated 24 Vdc (+10 %, -50 %) supply for input signal:
Internally generated.
Figure 8
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)
Remote I/O Jumpers Conn. D-Shell Male 25 Pin
Remote Interlock (Remote HV On/off)
Ch1 HV input: Pin 8, Pin 21
Ch2 HV input: Pin 9, Pin 22
Ch3 HV input: Pin 10, Pin 23
Technical Information
ION Pump Controller Description
24 Vdc insulated
External
Contact
5