4
Installation Procedure
Error Messages
Error Messages
On Front Panel
During the controller operation, if an error condition is detected, the
HV LEDs on the front panel will give the indication:
Led HV 1,2 flashing: error condition:
In Local-Remote I/O mode, the Reset of the Fault condition is
accomplished by switching to Off (0) the front panel switch of the
corresponding channel.
In Serial mode, the Reset of the Fault condition is accomplished by
sending the "F0n" command by the computer.
If one of the following conditions happens while the HV of the channel is On,
NOTE
the output of the channel is immediately set to Off, the LED on the front panel
indicates the corresponding Fault condition and the Fault information is sent to
the Serial output.
70/106
1 flash every 4 seconds = Interlock HV cable
2 flashes every 4 seconds = remote Interlock
3 flashes every 4 seconds = HV Over voltage (This Fault doesn' t
switch-off the HV)
4 flashes every 4 seconds = error inside ADC
5 flashes every 4 seconds = overcurrent on HV (short circuit)
6 flashes every 4 seconds = the Local/Remote switch was moved
with HV ON
7 flashes every 4 seconds = over temperature inside unit
8 flashes every 4 seconds = overcurrent in Protect (I>8 mA for
40W channel and I>16 mA for 80W channel)
9 flashes every 4 seconds = HV Under voltage
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)