SERIAL Operation

The SERIAL is set in order to operate by connecting the controller to an external
NOTE
PC that uses the "HYPER TERMINAL" or equivalent program (for example a
proprietary user program) running under Windows Operating System or other.
To operate in serial mode, connect the serial port to the controller "Serial"
connector on the rear panel, run the Windows "HYPER TERMINAL" program on
the PC with the following default settings:
This way of operation will allow to verify the functionality of the unit in the
"SERIAL" operating mode, give the commands and get all the operating data.
In the serial operation, depending on the user environment, the O.S. used must
NOTE
have real time capability.
See the "Technical Information" paragraph for details.

START/PROTECT Operation

The controller can operate in two modes: Start and Protect.
In Start, the 3 output channels can provide all the output power
regardless of the pump condition up to the short circuit condition.
This operating mode must be used to start the pump at high pressure.
In Protect, the controller limits the output current and switches off
the HV when the current exceeds the threshold value (8 mA) for more
than 2 seconds.
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)
COM1
9600 baud
No parity
8 data bits
1 stop bit