Enable, Disable Value in Floating Point Report Mode
(Access Level 1 Required).
"WR[space]Y[space]X"[Enter]
About Report
The continuous report sent by the IPCU protocol communication is
structured as follow:
At each main power on start up six remarking lines are sent.
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)
WR =
Write
Y =
31 (parameter number)
X =
0 for Standard Report Mode (default value)
1 for Floating Point Report Mode
The first line sent is the firmware release, for instance:
"* REL [20051117 ICPU]".
The second line is the notification that the EEPROM's
parameter are wrote into RAM:
* EVT [LOAD_EEPROM_PARAMETERS_INTO_RAM]
The third line is the notification that the calibration of the
Analog to Digital Convert used for the measurement of
output high voltage and current are done:
* EVT [ADC_CALIBRATION]
The fourth line is reporting that the internal I/O BUS are
initialized:
* EVT [IO_BUS_INIT]
The fifth line is the notification that a main power on cold
system startup reset is happened:
* EVT [COLD_RESET_SYSTEM_STARTUP]
The sixth and last line reports the number "n" of Power On Cold
Reset happened up to that moment:
* POR [n]