Technical Information

About Commands

About Commands

When the unit receives a command, it transmits the plus symbol: "+"
to confirm the command is executable.
If the command is not correct, the unit transmits the
minus symbol: "-".
When the parameter of a command is out of range, the unit sends:
"- [PARAMETER_ERROR]".
The unit receives continuously input characters from host but they
are hold only if the panel switch "Local-Remote/Serial" is on Serial.
If this switch is on Local-Remote and a character is received, the unit
sends:
"- [LOCAL_MODE]"
to inform that a character is received but it is not possible to execute
commands coming from the serial line.
When a command received is formally correct but can no be executed
because the status of the unit does not allow this, (for instance a
command: "A011" is sent but the cable interlock is not OK) the unit
sends:
"- [COMMAND_UNEXECUTABLE]"
to inform that the request is not coherent with the status of the unit.

Report TYPE 0...6

The User Operator Reports Type 0, 2, 4 and 6 display the 'continuous
average value' for the HV Channel Current, Voltage and Power, both
in normal or scientific notation and the 'instantaneous measured or
control values' in the hexadecimal format.
92/106
Ion Pumps Control Unit (No Battery) User Manual / 87-900-123-01 (B)