Information for the tester
Low-Resistance Measurement and Continuity Check (R)
Measuring method:
Current-voltage measurement as per
IEC 61557-4
Battery voltage: ≥ 5,5 V; max. 9 V DC
Nominal voltage:
≥ 200 mA DC
Short-circuit current:
Permissible overload:
U
= 600 V (measurement is not started)
rms
Measured-value formation
without zero compensation: R
= R
display
resistance)
Measured-value formation
with zero compensation
R
= R
display
3.10)
Compensation of measuring cables up to 5 Ω possible.
Measuring Range
Resolution
0,01...0,15...999 Ω
0,01-0,1 Ω
approx. 20 Ω
Internal resistance:
Measuring time:
approx. 2 s incl. voltage pole change for resistive
specimen
rising to max. 20 s for parallel inductances.
Permissible inductive
component:
max. 2.5 H without influence on measurement result;
max. 5 H without damage
Permissible series-mode
≤ 3 V AC or DC, no measurement is started beyond
voltage:
this value
Continuity check:
Same as low-resistance measurement, continuous measurement without pole
change after pressing the key. Response threshold: At Rext < 2 Ω ±0.5 Ω audible
signal and „con" is displayed.
Resistance Zero Compensation (<o>)
The measured value stored in this function is automatically subtracted in all
resistance measurements (R
, R
, R)!!
A
S
Measuring Range
Resolution
0,01-0,15-2,99-5,0 Ω
0,01-0,1 Ω
Measuring time:
approx. 1 sec; no voltage pole change
Battery voltage: ≥ 5.5 V; max. 9 V DC
Nominal voltage:
Short-circuit current:
≥ 200 mA DC, < 450 mA DC
- 70 mΩ (measuring-lead
measured
- R
( <o> / section
measured
compensation
Intrinsic error
± (5% of m.v. + 3 digits)
Intrinsic error
± (5% of m.v. + 3 digits)
59