Information for the tester
Measurement of Voltage
Measuring Range
Display Range
1 – 400 V
0 – 500 V
Intrinsic error:
± (1 % of m.v. + 1 digit) for sinusoidal signal
Internal resistance:
approx. 470 kΩ (U
Measuring sequence:
approx. 3 measurements/s
Permissible overload:
U
= 600 V AC
rms
Measurement of frequency
Measuring Range
Resolution
15,3-99,9-457 Hz
0,1-1 Hz
Intrinsic error:
± (0.1 % of m.v. + 1 digit)
Internal resistance:
approx. 470 kΩ (U
Measuring sequence:
approx. 3 measurements/s
Permissible overload:
U
= 600 V AC
rms
Voltages outside the specified frequency range (e.g. direct voltage) cannot
be measured.
Fault Current Circuit Breaker Test (RCD)
With/without tripping, with/without selective (x2), with/without probe, x5 and ramp function for
determination of the tripping current
Measuring method:
Lowering of voltage (without probe) or current-voltage
measurement (with probe) as per IEC 61557-6
Nominal voltage:
101 - 253 V sinusoidal
Frequency range:
46 - 65 Hz
1)
Mains impedance angle:
≤ 18 °; cos ϕ ≥ 0.95
Additional error
for cos ϕ < 0.95:
1)
+..% of m.v. = (1 - cos ϕ) x 100
Permissible overload:
U
= 600 V AC (measurement is not started)
rms
Electrical safety is given during and functional safety after overload.
Resolution
Frequency Range
1 V
46 – 65 Hz
: approx. 940 kΩ)
L-N
Dynamic Range
1-400 V (U L-N : 4-400 V)
: approx. 940kΩ)
L-N
51