Display Symbols, Error Messages - LEM SATURN 100 plus Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Measuring

3.11 Display Symbols, Error Messages

Function
Display
Meaning
before
Voltage cannot be
pressing
measured
„Start"
key
Overrange
No measurement
possible, frequency
outside permissible
range
No measurement
possible, battery
voltage too low
No measurement
possible, voltage
between sonde and PE
connection greater than
70 V
No measurement
possible, probe
connection not OK
after
pressing
„Start"
key
No measurement
possible, as voltage
outside permissible
range
Legend:
= flashing display
Notes
Frequency outside 46-65 Hz
or DC voltage or no voltage
Voltage greater than 500 V
Frequency greater than 457
Hz
Frequency less than 15,3 Hz
or DC voltage or voltage less
than 0,1 V
Replace batteries
Sonde connection carries
voltage greater than 70 V
with respect to PE
connection e.g. noise
voltage, mains voltage etc.
In position R A: : no probe
connected
In position R ISO: probe
connected
Voltage less than 101 V (FI,
R E , R S) possibly connection
missing
Voltage too high
FI, R A : voltage>253 V
R LOOP : voltage>435 V
R ISO :
voltage> 50 V
R:
voltage>3 V bzw.
current>50 mA
39