Troubleshooting - ProMinent DULCOTEST CLE 3-CAN-10 ppm Instructions Complémentaires

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4
Left LED
(Device LED)
Colour Flash code
red
illuminated
red
flashing*
red
slowly
flashing**
red
double
flashing***
green
illuminated
-
dark
*
**
Right LED
(CAN-open LED)
Colour Flash code
red
any
green
illuminated
green
flashing
Ignore the flash codes for approx. 2min (acknowledge alarm, if any) after
connecting the chlorine sensor.
If the LEDs repeatedly start to send one and the same sequence of flash
codes, the bus has to supply too many devices.
In this case, loop a (further) N module into the bus (see operating instructions
Multi-Channel Measuring and Control System DULCOMARIN
Pool Controller and Disinfection Controller DXCa Part 1, Assembly and Instal-
lation). In case of all other flash codes, contact the customer service!
®
ProMinent

Troubleshooting

Cause
Electronics error
Start-up phase
Calibration incorrect
0 ppm > measuring
value > 10 ppm
Measuring value <>
limit value
No correction
value pH
No device defect
present
No supply voltage
not functioning
***
Cause
Bus error
Bus status
OPERATIONAL
Bus status
PRE-OPERATIONAL
Result
Sensor faulty
No measuring value
communication
Measuring value
incorrect
Measuring value
too high / too low
Violation of limit
value
Correction value
pH missing
Standard operation
chlorine sensor
Chlorine sensor
connections
Result
No measuring value
communication
Standard operation
bus
Presently no
measuring value
communication
Troubleshooting
Remedy
Return chlorine
sensor or contact
customer service
Wait briefly
Re-calibrate
Check concentration
in the sample water
with DPD
Clarify cause; reset
values, if required
Check parameters
and configuration.
Check pH sensor
-
Check cable
Remedy
Contact customer
service
-
Wait briefly
®
II Swimming
13