Detection of equipment
7.2 Operate scales in a wireless group (menu)
......
rF
Autosend
......
58 •
The channels must be sufficiently far apart to ensure
troublefree data transmission. We recommend a spac-
ing of at least 30. Each channel number may only be
used for one channel.
Example of configuration; channel numbers when
setting up 3 wireless groups within one surgery:
• Wireless group 0: C1=_0, C2= 30, C3=60
• Wireless group 1: C1=10, C2=40, C3=70
• Wireless group 2: C1=20, C2=50, C3=80
If you wish to set up a wireless group with the scales,
devices
they will search for other active devices from the
seca 360° wireless
shown as modules in the display on the scales (e.g.
MO 3). The meaning of the numbers is as follows:
• 1: Personal scales
• 2: Length measuring device
• 3: Wireless printer
• 4: PC with seca USB wireless module
• 7: Baby scales
• 5, 6 and 8-12: Reserved for system expansion
All functions required to operate the device in a
seca wireless group can be found in the submenu "rF".
Information how to navigate in the menu can be found
on page 52.
Learn
Group (ID)
•
On
•
Off
•
On
System
•
Off
•
HI
•
MA
Autoprint
•
HI_MA
•
Off
•
Year
•
Month
•
Day
Time
•
Hour
•
Minute
system. The devices detected are
•
Channel 1 (C1)
Stop
•
Channel 2 (C2)
•
Channel 3 (C3)
Reg. Devices (MO)