Seca 704 Mode D'emploi page 64

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 71
Malfunction
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
... only the item "SYS" is
visible in the rF menu?
... only the items "SYS"
and "Lrn" are visible in
the rF menu?
... the items "APrt" and
"tIME" are not visible in
the rF menu?
Er:X:11
...
is displayed?
Er:X:12
...
is displayed?
Er:X:16
...
is displayed?
... when the Enter key
(send/print) is pressed
Er:X:71
and
played?
64 •
• The device was unable to send measured
results to the radio receiver (seca wireless prin-
ter or PC with seca USB wireless module).
• Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
The wireless module is disabled.
- Activate wireless module (see "Activate wire-
The wireless module is activated but no wireless
group set up.
- Set up wireless group (see "Set up wireless
No wireless printer is logged into the wireless
group.
- Use menu item "Lrn" to log the wireless prin-
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight
- Restart scales
The scales have been switched on with too great
a load.
- Reduce load
- Restart scales
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Restart scales
Data transmission not possible, wireless module is
disabled.
is dis-
- Activate wireless module (see "Activate wire-
Cause/Remedy
– Check that the scales are integrated in
the wireless network.
– Check that the receiver is switched on.
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.
less module (SYS)" on page 59).
group (Lrn)" on page 59).
ter into the wireless group(see "Set up wire-
less group (Lrn)" on page 59).
more evenly
less module (SYS)" on page 59).

Publicité

Table des Matières
loading

Table des Matières