M.blue
®
Fig. 22: Pressure in (cmH O) flow rate in (ml/h)
Pressure-flow characteristics of selected pressure lev-
els of M.blue in the vertical body position:
a) M.blue 0 b) M.blue 5 c) M.blue 10 d) M.blue 15
PRECAUTIONARY MEASURES AND
CONTRAINDICATIONS
FUNCTIONAL SAFETY AND
COMPATIBILITY WITH DIAGNOSTIC
PROCEDURES
WARNING
If a magnetic field is being applied and pres-
sure is applied to the valve at the same
time, thus triggering the brake mechanism,
it is not possible to rule out valve adjust-
ment. In MRI imaging M.blue creates arte-
facts which are larger than the valve itself.
CAUTION
For people using cardiac pacemakers: It
is possible that the function of the pace-
maker is influenced by the implantation of
a M.blue.
ADVERSE REACTIONS AND
INTERACTIONS
EN