Removal of transport stabilizers
On the bottom of the unit are four M6
stabilizing screws that must be removed
before start-up. They are marked with a
red warning washer.
Please retain the stabilizing screws for
future transport of the compressor.
Wheel installation
Wheels must be installed following the figure below for versions of the
compressor with wheels.
Compressed air connection
The medical compressor is equipped with quick couplings WALL (10) and
OUT (9) on the rear of the cabinet.
Connect the pressure hose from the given equipment/respiration equipment
to the quick coupling OUT (9) -outlet compressed air.
Applicable to compressors with WALL connection:
Connect the compressed air line from the central distribution line o the quick
coupling WALL (10) -inlet compressed air. Air from the distribution system is
automatically connected through the compressor to the OUT outlet port. In
this configuration, the compressor serves as a backup source of compressed
air. If the air pressure from the central distribution system is reduced, the
compressor automatically switches on and there is no interruption in the
supply of continuously pressurized air at the outlet of the compressor.
Air delivery from the compressor to the central air distribution pipeline is
prevented by non-return valve.
Please note that the air supply from the central distribution line
entering the compressor must be medical grade air (particulate size,
humidity.) The compressor does not modify the air from the central
distribution line.
The hose connecting the compressor to the respiration equipment
must not pass through a cold environment i.e. placed on the ground. It
should be as short as possible with no kinks (this may cause water to
condense inside the hose).
NP-DK50 DS-28_02-2021
13