EQUIPMENT DESCRIPTION
If the air pressure in central distribution drops below compressor outlet air pressure, the
controller automatically starts up the compressor, which delivers air directly to ventilator.
Air delivery from the compressor to the central air distribution pipeline is prevented by non-
return valve.
Primary ventilator compressed air source
The compressor continuously supplies compressed air directly to the medical ventilator
when the central compressed air distribution line is not connected to compressor air inlet
(WALL).
If air consumption is zero, the device switches to STANDBY mode.
The compressor must not be used for supplying air to the medical air
central distribution pipeline system. Do not connect compressor air
outlet to the central distribution pipeline system.