displayed. The display shows
"rEC X"
to be recalled.
recalled. The indicator will change to the recalled limits, weighing units and settings for the beeper and bar
graph.
NOTE:
If the recalled limit is for parts counting or percent weighing, the display will show the last sample
value used, ready for a new sample to be counted. If the memory location was empty the scale will return to
weighing.
2.7
PERCENT WEIGHING
The scale will use a mass on the platform as the 100% reference weight or input a reference weight using the
keypad.
Steps:
Press [Func]. The first option is
displayed. Press [Tare] again to enter percent weighing. Put the sample on the
platform. The scale will set the sample mass on the platform as 100% reference
weight.
NOTE:
If there is no reference weight on the pan and percent weighing function is
entered, pressing [Tare] again will return to normal weighing.
Remove the sample weight. Then any other weight placed on the platform will be displayed as a percentage
of the original sample.
The number of decimal points will depend on the weight used. A smaller weight will show only "100%" while a
larger weight might show "100.00%".
If the scale is showing zero When entering this function, Users can also input the reference weight using the
keypad, press [Tare] to accept the reference weight. The display will show "0.00 %".
"x x . x x %"
If the indicator shows
the numeric keypad to enter a new value. Press [Tare] to accept the new reference weight.
Press [Unit] to return to normal weighing.
NOTE:
1) The weight entered must be greater than 50 scale divisions.
2) The display may jump by large numbers unexpectedly if small weights are used to set as 100%
reference. For example, if only 23.5g is placed on a scale with 0.5g increments and is set to 100%,
the display will show 100.00%. However a small change of weight will cause the display to jump to
102.13% as an increase of one scale division (0.5g) to 24.0g will be equivalent to an increase of
2.13%.
© Adam Equipment Company 2017
"rEC"
. Enter the number corresponding to the desired memory location (0 to 9)
will be displayed for 2 seconds indicating the values stored in the location "X" is being
"FUnC 1"
, Press the [Tare] key.
, which is the last weight used as a reference, press [CE] to clear and use
"F1 PCt"
5
will be