2.6 CHECK COUNTING
Check-weighing is a procedure to display an indicator or sound an alarm when the weight on the platform
meets or exceeds the values stored in the memory. The memory holds values for a high limit and a low limit.
Either or both the limits can be set by the user.
The alarm and the LED bar graph can each be set to OFF. The LCD display will indicate whenever the
weight is within or exceeds the limits by showing
"LO"
high limit) or
( mass is below the low limit).
The limits can be locked by the manager (see the menu structure section) .A Limit Password must be used
to change the limits or operation of the beeper or bar graph.
Steps:
1) In normal weighing, Press the [Low Limit] key. It will show the current low limit. The
appear on the display.
2) Press the [CE] key to clear the old value and then enter the new low limit using the numeric keys.
Then press [Tare] to accept the value. If you want to reset the value to zero, press [CE] to clear the
value. The limits are displayed in the weighing unit in use. If the weighing unit is pounds:ounces, the
limits are entered in pounds and decimal parts of pounds. i.e. 6,0125 lb.
3) To set the high limit press [High Limit], the
the low limit was set. Pressing the [Tare] key to enter the value will return the scale to weighing, with
the Check-weighing function enabled.
4) During parts counting and percent weighing, the limits are set in the same way as above. The limits
are displayed in pcs or %.
5) To disable the check weighing function, enter zero into both the limits as described above. When the
current limits are shown, press [CE] to clear the settings, then press [Tare] to store the zero values.
NOTE:
The weight must be greater than 20 scale divisions for the check-weighing to operate.
Limits storing and recalling
The indicator can store up to 10 sets of high and low limits in memory along with the weighing units in use
(including pcs and %) as well as settings for the beeper and bar graph. During Check weighing the current
limits can be stored or previously stored units can be recalled.
If you are already in the check weighing mode the display will ask if you wish to
store the current limits by showing
"rECALL"
showing
. The [
"rECALL"
.
If you want to store the limits, when
"St ".
The display shows
"St X"
location (0 to 9).
where the current limits, weighing units and settings for the beeper and bar
graph are stored. The indicator will continue to work with the current settings as
active.
If you wish to recall any of the pre-stored limits, press [Tare] when
© Adam Equipment Company 2017
"StOrE"
or recall another set of limits by
Lim] key can be used to toggle between
"StOrE"
is displayed press the [Tare] key.
Enter a number corresponding to the desired memory
will be displayed for 2 seconds indicating the location X
"OK"
(mass is between the limits),
"HI"
symbol will be on. Set the high limit in the same way
"StOrE"
"rECALL"
4
"HI"
(mass is above the
"LO"
symbol will
and
is