C. If milk accidentally gets into the motor:
Turn off the pump and unplug from the power source. Use a damp
(not wet) cloth to wipe the motor and proceed as follows:
1. Remove battery compartment cover from bottom of motor.
2. Remove small, black foam piece from
bottom of pump.
3. Rinse the foam piece with clear lukewarm
water, and squeeze out excess water.
Allow to dry thoroughly before inserting
back into pump.
4. Pour about 2 ounces of warm soapy water
into a container.
5. Connect the tubing to the motor, leaving the
other end unconnected.
6. Connect the power cord to the motor, then
plug into electrical outlet (do not power with
battery).
7. Hold the end of the tubing that is not connected
to the pump in the container with the soapy water.
8. Place the motor over a dry sink and switch on. The water will
pump through the tubing into the motor and flow out through the
opening beneath the battery cover. Allow it to pump until all the
water has been flushed through.
9. Repeat this procedure again with 2 ounces of warm clear water.
10. Leave the motor switched on for an additional 3 minutes so it
dries completely.
11. As soon as the foam piece is dry, put it in place again, and
re-install the battery compartment cover.
12. If necessary, clean the pump kit according to cleaning
instructions.
D. Other troubleshooting:
1. The pump is designed to automatically turn off 30 minutes after last
setting adjustment.
2. When traveling outside the U.S., use battery power only. If you
experience discomfort while pumping, contact a
breastfeeding specialist.