Companion 5
Guidance and Manufacturer's Declaration± Electromagnetic Immunity
The Companion 5 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Companion 5 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic
environment – guid-
±8 kV contact
ance
±15 kV air
IEC 61000-4-2
±2 kV for power supply
lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/
output lines
±2 kV common mode on
AC lines
Surge
±1 kV differential on AC
lines
IEC 61000-4-5
±2 kV common mode on
outdoor I/O lines
0% U
45°, 90°, 135°, 180°,
Voltage dips, short
225°, 270°, 315°)
interruptions and
0% U
voltage variations
on power supply
70% U
input lines IEC
for 25/30 cycles (0°)
61000-4-11
0% U
(0°)
Power frequency
(50/60 Hz) magnetic
30 A / m
field IEC 61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
20 - ENG PN 21420526-C4 B | User Manual
Compliance level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
N/A
N/A
±1 kV differential on
AC lines
N/A
0% U
for 0.5 cycles (0°,
(0°, 45°, 90°, 135°,
T
180°, 225°, 270°,
315°)
for 1 cycle (0°)
0% U
T
(30% dip in U
)
70% U
T
T
U
) for 25/30 cycles
T
(0°)
for 250/300 cycles
T
0% U
cycles (0°)
30 A / m
Electromagnetic environment ±
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
for 0.5 cycles
T
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the Companion 5 re-
for 1 cycle (0°)
quires continued operation during power
T
mains interruptions, it is recommended
(30% dip in
T
that the Companion 5 is powered from
an uninterruptible power supply (UPS) or
a battery.
for 250/300
T
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.