HELiOS
Cleaning Standard
WARNING: CLEAN ONLY AFTER THE UNIT
IS EMPTY.
• Clean using a solution of mild dish washing
detergent and water.
• Apply cleaning solution directly to a lint-free
cloth. Approved cleaners include HydroPure
and HydroKlean. Do not spray cleaners
directly on the HELiOS Reservoir.
• Wipe the outside surface with the lint-free
cloth until the outside surface is clean.
Caution: Do not use high temperature
and high pressure washing equipment
to clean these units.
• Do not get cleaner on any internal compo-
nents or valves.
• Allow the unit to dry thoroughly before using.
Note: Note to health care provider – for reprocessing
procedures, see applicable service manual.
Disposal
Always return HELiOS Reservoir, including all
components, to your homecare provider for proper
disposal. You can also contact your local city or
town offices for instructions on proper disposal of
the battery.
WEEE and RoHS
This symbol is to remind the equipment
owners to return it to a recycling facility at
the end of its life, per Waste Electrical and
Electronic Equipment (WEEE) Directive.
Our products will comply with the restriction of
Hazardous Substances (RoHS) directive. They will
not contain more than trace amounts of lead or
other hazardous material content.
Transport and Storage
The device should be stored in the upright position,
and be well ventilated. Do not allow the device to
lie on its side. Humidity up to 95% noncondensing.
Temperatures range from -40°F to 158°F (-40°C to
70°C).
Operating temperature ranges from 14°F to 104°F
(10°C to 40°C). Relative humidity range from 30%
to 75% noncondensing.
Note: The atmospheric pressure range is 700 hPa to
1060 hPa (elevation of 10,000 Ft. to -1,000 Ft.).
10 - ENG
PN 21157307-C4 B | User Manual
Accessories
15 in (38.1 cm) Roller Base for
HELiOS Reservoirs, available for
all models.
NOTE: Only use roller base on flat
surfaces.
0-10 LPM External Flow Control
Valve
Allows you to breath continuously
from it between 0-10 LPM. If pre-
scribed, your health care provider
will provide it.
50 ft (15.24 m) HELiOS Oxygen
Supply Line
Used to connect an H300 or H850
portable unit to your reservoir to
allow the portable to be used as a
conserving device.
NOTE: When using the HELIOS
oxygen supply line with the H850,
the portable must be in demand
mode.
Note: To Equipment Provider: The following oxygen
administration accessories are recommended for use
with the HELiOS:
• Nasal Cannula: CAIRE Part Number 6-778057-00
• Firebreak: CAIRE Part Number 21126636
A firebreak is recommended for use with any can-
nula.
• CAIRE offers a firebreak intended to be used in
conjunction with the oxygen reservoir. The firebreak
is a thermal fuse to stop the flow of gas in the event
that the downstream cannula or oxygen tubing is
ignited and burns to the firebreak. It is placed in-line
with the nasal cannula or oxygen tubing between
the patient and the oxygen outlet of the HELiOS.
For proper use of the firebreak, always refer to the
manufacturer's instructions (included with each
firebreak kit).
• For any additional recommended accessories,
please see the Accessories Catalog (PN MLLOX0010)
available on www.caireinc.com.