Apex iX Sense CPAP System Manuel D'instructions page 25

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 109
iX Sense / English
Air is leaking from
around mask or tubing
or device
(Low Pressure)
Screen :
show flashing icon
Therapy data has not
been sent to my care
provider
Caution :
1.
Relevant risks need to be analyzed and evaluated by manufacturer. Relevant protective measures should
be taken.
2.
Any changes to the relevant network configuration and network device connection methods may cause
unknown risks. If necessary, please contact your dealer or manufacturer.
1.
Tubing is damaged.
2.
Tubing is not correctly connected.
3.
Humidifier is not properly
assembled.
4.
Water chamber is not properly
assembled.
5.
Mask is not properly connected.
6.
User does not wear the mask
properly.
1.
Wireless coverage may be poor.
2.
The No wireless connection icon is
displayed on the top right of the
screen. No wireless network
available.
3.
Data transfer is not enabled for
your device.
1.
Check the condition of the tubing.
2.
Connect the tubing properly.
3. Connect the humidifier properly.
4.
Connect the water chamber properly.
5. Make sure the mask is perfectly sealed
on the user's face.
6. The device will automatically turn OFF if
Low Pressure Alarm persists over 3
minutes.
Screen : flashing icon will disappear
1.
Make sure that the device is placed
where there is coverage (ie, on your
bedside table, not in a drawer or on the
floor). The icon of wireless should be
turned on if the wireless coverage is
good.
2.
Disconnect the device from mains
supply, wait 5 seconds, then reconnect.
The device will automatically re-upload
data.
3.
Make sure there are no co-existence
interfere by adding new item/dis
connect item to the connected
IT-network.
4.
Talk to your care provider about your
settings
22

Publicité

Table des Matières
loading

Table des Matières