4.1.4. Isolation (IEC 61557-2, 1997)

Range
Resolution
Accuracy
Measuring current
Max. direct current
Nominal voltage
Voltage at open circuit 500 V
Protection

4.1.5. Current (with MN73 clamp)

Rating
Resolution
Accuracy
Input resistance
Protection
(∗) typical
The DC component is cut by a capacitor at the tester input.

4.1.6. Voltage

Rating
Resolution
Accuracy
Input impedance
Protection
The DC component is cut by a capacitor at the tester input.

4.1.7. RCD test

4.1.7.1. Range for use
Installation voltage
Frequency
Max. internal temperature
44
www.electronicaembajadores.com
0.1 ... 0.5 ... 200.0 MΩ
100 kΩ
from 0.1 MΩ to 0.5 MΩ
:
from 0.5 MΩ to 0.6 MΩ
:
from 0.7 MΩ to 0.8 MΩ
: 3 %
from 0.9 MΩ to 10.0 MΩ : 3 %
from 10.1 MΩ to 200 MΩ : 3 %
1.0 mA
for R
500 kΩ
DC
4.0 mA
DC
500 V
DC
V
550 V
DC
480 Vrms
with MN73 clamp
0 to 200 A
0 to 2000 mA
100 mA
1 mA
± 2 (∗) % ± 7 D
± 1 (∗) % ± 7 D
1 MΩ
1 MΩ
480 Vrms
480 Vrms
0 - 600 V
1 V
± 1,5 % ± 5 D
1 MΩ
> 480 Vrms
190 to 290 V
45 to 65 Hz
80° C
Chapter II
-
± 1.5 D
(out of IEC 61557)
-
± 1.5 D
± 2.0 D
± 2.5 D
± 3.0 D
Voltage input of
tester alone
0 to 2000 mV
1 mV
± 1 % ± 5 D
1 MΩ
480 Vrms
Digital Earth Tester