Symbols; Instructions - Metrix MX 435D Notice De Fonctionnement

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chapter II
Do not perform earth, continuity or insulation resistance measurements on live circuits.
In the event of incorrect handling (accidental connection to an external voltage during
insulation, continuity or earth measurements), electrical protection systems are activated
(CTP). You must wait for them to cool down (≈ 3 minutes) before performing more
measurements.
In RCD mode, a high number of successive measurements may cause any new test
using thermal protection to stop momentarily. Users should then wait for sufficient
internal cooling before starting a new series of measurements.
If the symbol "BAT" is displayed, it is imperative to reload the accumulator pack or to
change the batteries.

1.1.3. Symbols

The following symbols, represented on the front panel, have the following meanings:
WARNING
DANGER
voltages.
Earth
Equipment protected throughout by double or reinforced insulation.
)
• • • • )
)
Buzzer
The rubbish bin with a line through it means that in the European Union, the
product must undergo selective disposal for the recycling of electric and electronic
material, in compliance with Directive WEEE 2002/96/EC.
Clamp
Conform to CE standard
Power supplied by an accumulator pack - 8.4V NiMH (CA model ref. HX0086 only)
or by two 9V batteries
This product is 100% recyclable.

1.1.4. Instructions

Before opening the device to access accumulator pack or the batteries, disconnect the
measuring circuit leads, then the device.
Adjustments, maintenance or repair work on the tester must only be carried out by
qualified personnel approved by the manufacturer.
In the event of faults or abnormal constraints, power down the instrument and do not
allow anyone to use it until it has been checked.
It is advisable to remove the accumulator pack or the batteries if the instrument is not
going to be used for a long time.
Digital Earth Tester
www.electronicaembajadores.com
Risk of DANGER ! Refer to the user's manual.
Failure to perform the instructions in this operating manual
preceded by this symbol, or to perform them correctly, may cause
bodily injury or damage to the instrument and the installations.
Risk of electrical shock
Withdrawal authorized on conductors containing dangerous
27