A
Parameter Values
The following table lists each of the setup parameters shown in the Setup mode. A descriptive name
is included, followed by a list of the selections with corresponding selection values.
Default values are indicated by an asterisk [*].
F1.1.
F1.2
F1.3
F1.4
F1.5
30094013
|
00
|
0 /2013
F-Code
Name
Scale Type
F1.1.1
Type
F1.1.2
Approval
F1.1.3
Remote Protocol
Capacity and Increment
F1.2.1
Unit
F1.2.2
Range
F1.2.3
Range 1 Capacity
F1.2.4
Increment Size 1
Range 2 Capacity
F1.2.5
[only available with F1.2.2=
2r]
Increment Size 2
F1.2.6
[only available with F1.2.2=
2r]
Calibration
F1.3.1
GEO
F1.3.2
Calibration
F1.3.3
Linearity Calibration
Zero
F1.4.1
Auto Zero maintenance
F1.4.2
Zero Blanking
F1.4.3
Power up Zero [Unit : %]
F1.4.4
Pushbutton Zero [Unit : %]
Tare
F1.5.1
Pushbutton Tare
F1.5.2
Tare Interlock
F1.5.3
Auto Tare
Auto Tare Threshold
F1.5.4
[only available with
Selections and Values
F1 Scale
*SCL
RCOM1
*None
OIML
NTEP
*Toledo Continuous, SICS
g
*kg
*1r: One range, 2r: Tow ranges
*50, Range :1~500'000
*0.01, Selections are based on resolution range: 1000~30'000
*50, Range :1~500'000
0.0001
0.005
0.0002
*0.01
0.0005
0.02
0.001
0.05
0.002
0.1
*16 [01 – 31]
Off
*0.5d
1d
*Off
20d
Off
2
Off
*2
Off, *On
*Off, On
*Off, On
Range 0~FS [Full scale] *10
METTLER TOLEDO IND231/IND236 User's Guide
RCOM2
AR
SRI
oz
lb
0.2
10
0.5
20
1
50
2
100
5
200
3d
10d
20dZero
*10
20
10
20
A-1

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ind236

Table des Matières