Clinical Restrictions; Maintenance,Transportation And Storage; Cleaning And Disinfecting; Maintenance - Gima OXY-200 Mode D'emploi

Masquer les pouces Voir aussi pour OXY-200:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

H Testee can not use enamel or other makeup.
I Please clean and disinfect the device after operating according to the User Manual (7.1).

6.3 Clinical Restrictions

A As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of
subject is required. For a subject with weak pulse due to shock, low ambient/body temperature,
major bleeding, or use of vascular contracting drug, the SpO
decrease. In this case, the measure will be more sensitive to interference.
B For those with a substantial amount of staining dilution drug (such as methylene blue, indigo
green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb)
or thiosalicylic hemoglobin, and some with icterus problem, the SpO
device may be inaccurate.
C The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major
factor resulted in serious error of SpO
D As the SpO
value serves as a reference value for judgment of anemic anoxia and toxic anoxia,
2
some patients with serious anemia may also report good SpO

7 Maintenance,Transportation and Storage

7.1 Cleaning and Disinfecting

Using medical alcohol to disinfect the device, nature dry or clean it with clean soft cloth.

7.2 Maintenance

A Please clean and disinfect the device before use according to the User Manual (7.1).
B Please recharge the battery when the screen shows low power alarm information.
C Recharge the battery soon after the over-discharge. The device should be recharged every six
months when it is not regular used. It can extend the battery life following this guidance.
D Users are advised to calibrate the device termly (or according to the calibrating program of
hospital). It also can be performed at the state-appointed agent or just contact us for
calibration.

7.3 Transportation and Storage

A The packed device can be transported by ordinary conveyance or according to transport
contract.The device can not be transported mixed with toxic, harmful, corrosive material.
B The packed device should be stored in room with no corrosive gases and good ventilation.
Temperature: -40°C~60°C; Relative Humidity: ≤95%
measure.
2
16
waveform (PLETH) will
2
determination by this
2
measure.
2