Guidance and manufacturer's declaration – electromagnetic immunity –
Guidance and manufacturer's declaration – electromagnetic immunity
The CMS70A is intended for use in the electromagnetic environment specified below. The
customer or the user of CMS70A should assure that it is used in such an environment.
Immunity
IEC 60601 test
test
±6 kV contact
Electrostatic
±8 kV air
discharge
(ESD)
IEC
61000-4-2
±2 kV for power
Electrical fast
transient/burst
supply lines
IEC
61000-4-4
±1 kV
Surge
IEC
differential mode
±2 kV common
61000-4-5
mode
Voltage dips,
<5% U
short
(>95% dip in
interruptions
U
T
and
voltage
for 0.5 cycle
variations on
power supply
40% U
input lines
(60% dip in U
IEC
for 5 cycles
61000-4-11
70% U
for all EQUIPMENT and SYSTEMS
Compliance
level
±6 kV contact
±8 kV air
±2kV for
power supply
lines
±1 kV
differential
mode
±2 kV common
mode
<5% U
T
(>95% dip in
)
U
T
for 0.5 cycle
40% U
T
)
(60% dip in
T
U
T
for 5 cycles
T
Electromagnetic environment -
level
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
T
that of a typical commercial or
)
hospital environment.
T
)
23
guidance