OPERATION OF THE DEMINERALIZER

Once installed and fitted correctly, the demineralizer is able to supply demineralized
water immediately. This water is ready to be used or to manually fill the sterilizer's tank
(see the sterilizer user manual).
ATTENTION
The demineralizer is equipped with a Water Block
flow after a pre-adjusted volume (5 liters) in order to prevent a large spillage of
water in case of a broken tube. When you take water from the demineralizer with
the use of the spray gun and the volume goes beyond 5 liters, the Water Block
valve will cut off the water supply. To reset the Water Block
"Maintenance instructions for the safety valve".
To avoid a water cutoff, release the water gun once every 30 seconds if quantities
of more than 5 liters are taken from the demineralizer.
WARNING
The Water Block serves as an anti-flooding device. Blocking only takes place if
there is considerable leakage (such as a cut in the tube between the Water Block
and the demineralizer). Blocking does not take place if there is a small leak in the
product's hydraulic system.
WARNING
It is recommended that the connection tap
between the spray gun's spiral tube and the Dem
10 be closed if the gun is not being used.
The quantity of demineralized water produced per set of resin cartridges depends on the
quality and hardness of the city water in your area. Refer to the table below for further
details:
300
250
200
150
100
50
0
12
0
200
400
Conductivity of the tap water ( µS/cm )
®
valve that cuts off the water
600
800
1000
®
®
valve see chapter
1200
Dem 10