Télécharger Imprimer la page

Diastolic Blood Pressure Reading; Recording Your Readings; Other Possible Malfunctions And Their Elimination - Microlife BP AG1-30 Mode D'emploi

Masquer les pouces Voir aussi pour BP AG1-30:

Publicité

Les langues disponibles

Les langues disponibles

4.4.4. Diastolic blood pressure reading

Allow the pressure to continue dropping at the same deflation rate. When your diastolic blood
pressure value reached, the thumping sound stops. Deflate the cuff valve completely. Remove
the cuff from arm and stethoscope from ears.
4.4.5. Record your readings
Repeat the measurement at least two times. Do not forget to record your readings and the time
of the day measurement is made immediately after you finish measuring. A suitable time is first
thing in the morning, or just before evening meals. Remember that your physician is the only
person qualified to analyze your blood pressure.
☞ Further information
Measurements should not occur soon after each other, since otherwise, the results will be
falsified. Wait therefore for 2 minutes in a relaxed position, sitting or lying, before you repeat
a measurement.

5. Other possible malfunctions and their elimination

If problems occur when using the device, the following points should be checked and if
necessary, the corresponding measures are to be taken:
Malfunction
The sound transmission is poor,
distorted or there is extraneous noise.
Remedy
1. Check the earpieces if they are plugged or cracked.
If not, make sure they do not fit poorly as worn.
2. Check the tube if it is broken or twisted.
3. Check the bell and diaphragm of chestpiece if there
is any crack.
4. Make sure the chestpiece is in proper contact with
skin and over brachial artery during measuring.
Clean or replace any defective parts if found to
avoid inaccurate reading.
9

Publicité

loading