6
Function check
6.1 Intervals
6.2 Performing a function check
Result
Carry out a function check at regular intervals:
•
After each hygienic preparation
•
After each repair
•
At least every 6 months
1. Check whether the module is set up and connected correctly in
accordance with the instructions for use.
2. Check the housing for cracks, damage, and heavy soiling.
3. Connect the module up to the device
the prismaCONNECT module", page
4. Check whether the
on the therapy device's display.
5. Connect the Ethernet cable up to the LAN port and connect it
to a PC or IT network.
6. Check whether the
therapy device's display.
7. If the module does not work properly or shows signs of
damage: Contact your authorized dealer.
The function check is complete.
(see "4.1 Connecting up
34).
symbol for prismaCONNECT is shown
symbol for network is shown on the
prismaCONNECT
6 Function check
EN
41