4.3 Configuring the prismaCONNECT
module
4.3.1 prismaLINE (WM 100 TD)
Requirement
•
•
1. Select the network settings on the therapy device: Expert sector
2. If required, select Type (default setting is DHCP).
Result
prismaCONNECT is configured.
The settings are saved in the prismaCONNECT module.
4.3.2 prismaVENT (WM 110 TD, WM 120 TD)
Requirement
•
•
1. Select the network settings on the therapy device: Expert sector
2. Select DHCP On or Off.
The prismaCONNECT module is connected
Connecting up the prismaCONNECT module", page
The therapy device is switched on and in Standby mode (see
instructions for use of the therapy device).
| Main Menu | Device | Network (see the instructions
Additional information for hospital personnel).
DHCP: The prismaCONNECT module is assigned a dynamic IP
address. If there is no DHCP server available, the IP address is
assigned via zeroconf/Auto-IP in the range from 169.254.1.0
to 169.254.254.255.
Manual: Configure the static IP address and a subnet mask.
It is also optionally possible to set a gateway and DNS.
The prismaCONNECT module is connected
Connecting up the prismaCONNECT module", page
The therapy device is switched on and in Standby mode (see
instructions for use of the therapy device).
| System | Device settings | Network settings
DHCP On: The prismaCONNECT module is assigned a dynamic
IP address. If there is no DHCP server available, the IP address
is assigned via zeroconf/Auto-IP in the range from 169.254.1.0
to 169.254.254.255.
4 Preparation and operation
(see "4.1
(see "4.1
prismaCONNECT
34).
34).
EN