4. Troubleshooting
You notice
The function symbol on the tracing grid has
been touched but the unit display has not
changed.
Area measurement does not appear in unit
display. Unit double BEEP's during/after tracing
is complete
and unit
0.0
display
flashes
previous area
reading.
No length or
- - - -
width results
available.
The display
screen shows
the cursor
moving
across the
screen.
The battery symbol flashes on and off and
subsequently
0.0
remains on
permanently.
The unit
double BEEPs
The unit will not switch 'ON' and the display
remains blank.
5. Maintenance and Storage
Replacing the Stylus Nib
• Insert the nib of the stylus into the gap of the metal ring.
• Squeeze the metal ring to secure the nib and pull to remove.
• Place a new nib directly into the tip of the stylus and press firmly to secure.
Storage
• Ensure the VISITRAK Digital stylus is always returned to the unit recess
when not in use.
Environmental Conditions
Operating temperature range:
Storage temperature range:
Operating humidity:
Storage humidity:
Cleaning Procedure
Clean VISITRAK Digital, stylus and coversheet in accordance with standard infection control
procedures, or as below:
1. Wipe down all device surfaces using a damp soft cloth and a low level
disinfecting solution.
2. Dampen another soft cloth with clean water and use this to wipe down all
surfaces to remove the disinfecting solution.
3. Dry with a separate soft cloth.
Do not immerse any part of VISITRAK or use a dripping wet cloth for cleaning.
Do not sterilize the VISITRAK Digital or stylus, damage or destruction may result.
Do not use plastic solvents or abrasive cleaners.
Caution
Care should be taken to ensure that moisture does not seep into the electrical components
of the unit or stylus.
Possible causes
Function symbol touched instead of the box
below the function symbol.
Incomplete tracing on digital tablet, which may
be due to:
• The side switch on the stylus has not been
pressed close enough to the unit surface.
• Tracing did not finish in the same place as it
was started.
2
• The stylus has been lifted during tracing or
cm
some point of the line has been repeated.
• The stylus/tracing has overlapped the grid area.
• The grid slipped on the VISITRAK device surface
during tracing of the wound outline.
Multiple tracing grids have been added together.
The length and width parameters cannot be
displayed following these functions.
The digital unit has an automatic power save
(sleep) function if not touched after 3 mins with
the stylus.
The batteries have reached a level which is
approaching unacceptable for the unit to function.
2
cm
Battery power is so low that there is not enough
energy to power up the unit.
15 - 25˚C (59 - 77˚F)
5 - 25˚C (41 - 77˚F)
10 - 85% RH (noncondensing)
10 - 95% RH (noncondensing)
Solution
Touch the box underneath the appropriate function
symbol on the tracing grid using the stylus.
Do not press the side switch on the stylus.
Fit the holes in the tracing grid over the pegs on VISITRAK
Digital. If available, place the coversheet on top of the
tracing grid and secure over the pegs on VISITRAK Digital.
Hold the coversheet/tracing grid on VISITRAK Digital
surface using the unoccupied hand. Press the stylus side
switch in close proximity to the tracing grid, and place
the stylus at the start point of the tracing to activate the
tracking symbol in the display. Trace the wound outline
on the grid, whilst ensuring that minimum pressure is
applied to the grid by the stylus. Do not lift the stylus
while tracing and remain within the grid area. Start and
finish the tracing at the same point. You will hear an
audible BEEP to indicate a successful trace.
For large wound areas requiring multiple tracing grids,
only the wound area can be measured.
Touch the tracing grid with the stylus and the unit will
reactivate. Proceed with the next function.
New batteries should be purchased and
inserted into the unit.
6. Other Information
Specifications
Physical Characteristics
Height
35.1mm
Length
300.7mm
Width
207.1mm
Weight
688g
Electrical Characteristics
Input
9 Volts DC
Power Source 6 x 1.5V Batteries (AA or LR6)
Classification and Safety
Internally powered, classified in accordance with
IEC 60601-1-1 and IEC 60601-1-2.
Faults and Technical Assistance
If your VISITRAK Digital unit develops a fault, or if there are signs
of damage, contact your local market representative for assistance.
Manufacturer
Made in the United Kingdom for
Smith & Nephew Medical Limited, Hull HU3 2BN, England.
™Trade Mark of Smith & Nephew.
© Smith & Nephew 2003
www.smith-nephew.com