NOTE:
The factory zero setting may not always be precise,
particularly after a probe tip exchange. For best accuracy, the
G age should be zeroed on the included glass plate.
Units
Converts the display from mils to microns and vice versa.
Reset
Reset (menu reset) restores factory settings and returns the
Gage to a known condition. The following occurs:
- All batches, stored measurements, batch names and screen
captures are erased.
- The Zero adjustment is returned to the factory setting.
Since this may not always be a precise zero, the Gage
should be checked on the glass plate.
- Menu settings are returned to the following:
Memory = OFF
Statistics = OFF
Hi Lo Alarm = OFF
Auto Dim = ON
Perform a more thorough Hard Reset as follows:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the ( + ) and
buttons until the Reset symbol
This returns the Gage to a known, "out-of-the-box" condition.
It performs the same function as a menu Reset with the addition of:
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:
Units = microns
Flip Display = Normal
Auto Sync .net = ON
NOTE:
- Keep the gage away from metal during a Reset.
- Date, Time and WiFi settings are not affected by either Reset.

Setup Menu

Bluetooth & Stream = OFF
WiFi & Access Point = OFF
USB Keyboard & Stream = OFF
Display = None
Sound = ON
Language = English
Backlight = Normal
4
appears.
Battery Type = Alkaline
Bluetooth Smart = OFF
USB Drive = ON